Reading an OTDR trace: events, dead zones, ghosts, gainers · Operations: splicing, testing, maintenance · Fibre-optic links · Documentation · CodingBox
CodingBox Documentation

Reading an OTDR trace: events, dead zones, ghosts, gainers

The OTDR draws the fibre as a line sloping down with distance and marks every place where light was lost or reflected. Reading that picture correctly — telling a splice from a bend, a ghost from a break, a real gain from a measurement artefact — is the core skill of fibre acceptance and fault location. This page explains what the instrument actually measures, how each setup parameter changes the trace, what every event type looks like and what its numbers mean, the dead-zone and dynamic-range limits, why bidirectional measurement is not optional, the standard misreadings, and how to turn a distance on the trace into a place on the route.

An OTDR trace: what each event looks like

How the OTDR sees the fibre

MechanismWhat happensWhat you see
Pulsea laser pulse of 5 ns to 20 µs is launched at 1310, 1550, 1625 or 1650 nmnothing directly — the pulse is the probe
Rayleigh backscattera tiny fixed fraction of the pulse (about −79 dB at 1310 nm, −81 … −82 dB at 1550 nm for a 1 ns pulse on G.652) scatters back from every point of the fibrethe sloping line; it falls at twice the fibre attenuation because the light travels out and back, and the instrument divides by two to display one-way loss
Fresnel reflectionat every index step (connector, air gap, break, end) a much larger fraction returnsspikes above the backscatter line
Time to distanced = c · t / (2 · n), with the group index n (IOR) of the fibre — 1.4675–1.4685 for G.652 at 1550 nm, from the fibre data sheetthe distance axis; an IOR error of 0.0015 shifts every distance by 0.1 % (10 m on 10 km)
Averagingthousands of pulses are averaged; noise falls with the square root of the counta smoother trace after 30 s to 3 min

Setup parameters

ParameterChoiceEffect on the trace
Wavelength1310 + 1550 nm for loss; 1550 vs 1310 comparison for bends; 1625/1650 nm for bend hunting and for live fibre through a filtersplices look the same at all wavelengths, bends get worse with wavelength (Attenuation & windows)
Pulse widthshort (5–30 ns) — best resolution and shortest dead zones, range under 5 km; medium (100–300 ns) — 10–40 km; long (1–20 µs) — 100–250 km with dead zones of hundreds of metresrule: the shortest pulse that still shows the far end 6 dB above the noise; take two traces — short for the near end, long for the range; auto modes do this
Range1.5–2 × the fibre lengththe end reflection and any ghosts stay on screen
Acquisition time30 s to 3 min; longer for long pulses and far endsnoise floor drops, small events become measurable
Index of refraction (IOR)the fibre data sheet value; the same value for the baseline and every later comparisondistances; never "tune" IOR to make the trace match the drawing
Backscatter coefficientdata sheet value for the wavelengthreflectance numbers only — loss is unaffected
Launch and receive cords150–1 000 m each, longer than the attenuation dead zone of the pulsethe first and last connectors of the link become measurable
Thresholdssplice 0.1 dB, connector 0.5 dB, reflectance −40 dB (UPC) or −55 dB (APC), end-of-fibre 3–5 dB, section loss 0.4/0.25 dB/kmthe event table flags what exceeds them
Live-fibre checkthe instrument warns if it sees traffic light at the portprotects the OTDR receiver and confirms the fibre is dark (Safety & handling)

Event catalogue

EventSignature on the traceTypical numbersNotes
Launch connectorlarge spike at 0 m followed by a dead zone−14 … −45 dBmeasure it through a launch cord; an APC launch cord gives a smaller spike
Fibre sectionstraight line sloping down0.33–0.35 dB/km at 1310, 0.19–0.22 at 1550, 0.20–0.24 at 1625 nma slope above 0.4 dB/km at 1550 means stress, water, microbending or old fibre — compare wavelengths
Fusion splicesmall step down, no spike, the same size at every wavelength0.02–0.10 dB; redo above 0.15 dBmay show as a gain in one direction (see gainer)
Mechanical splicesmall step plus a small spike0.1–0.5 dB, −40 … −55 dBreflectance grows as the gel ages
Connector pair (UPC)spike plus step0.2–0.5 dB, −45 … −60 dBa dirty pair: bigger step, spike toward −30 dB (Endface inspection & cleaning)
Connector pair (APC)step with little or no spike0.2–0.5 dB, below −60 dBlooks like a lossy splice — check the records
Macrobendstep down with no spike, much larger at 1550/1625 than at 1310 nm0.1–3 dB at 1550, near zero at 1310the wavelength dependence is the signature (Plant faults)
Crush, pinch, ice, tight tiessteeper slope over a section, wavelength dependent+0.1–1 dB/km locallyoften seasonal
Gainerapparent step up−0.05 … −0.3 dBa splice between fibres with different backscatter (mode field diameter, index); the true loss is the average of the two directions
Exaggerated lossthe mirror image of a gainer in the other direction+0.1 … +0.3 dB too muchsame cure
Ghosta spike with no loss, at twice (or a multiple of) the distance of a strong reflector, often beyond the fibre endreflectance-likemoves or vanishes when pulse, range or launch conditions change; reduce the reflection that causes it (Reflections & return loss)
End of fibre, flat or UPClarge spike, then a drop into the noise−14 … −30 dBthe classic end
End of fibre, APC or angled breaka drop into the noise with little or no spikesmall reflectanceset end detection by loss, not by reflection
Breakreflective or not, then noiseanya wet or crushed break can reflect almost nothing
Splitter (PON)large step with no spike1:2 = 3.5 dB, 1:8 = 10 dB, 1:32 = 17 dBneeds high dynamic range; PON-optimised OTDRs exist
Noise floorragged flat tail—nothing beyond roughly (dynamic range − 6 dB) is measurable
Saturationflat-topped spike—reflectance cannot be measured — use a shorter pulse or a launch cord
Water or hydrogenslope increase, worst at 1383 and 1625 nm—multi-wavelength comparison (Reliability & ageing)

Dead zones and dynamic range

Pulse widthEvent dead zoneAttenuation dead zoneTypical range on G.652 at 1550 nm
5–10 ns0.8–1.5 m3–5 m1–3 km
30 ns3 m8–10 m5 km
100 ns10 m15–25 m10–20 km
300 ns30 m40–60 m30 km
1 µs100 m150–250 m50–80 km
3 µs300 m500–700 m100 km
10–20 µs1–2 km1.5–3 km150–250 km

Values are indicative and instrument dependent. Event dead zone is the minimum distance between two reflective events that are still shown separately (the spike width 1.5 dB below its peak). Attenuation dead zone is the distance after a reflection before the trace returns to within 0.5 dB of the backscatter — a splice inside it is invisible. Dynamic range is the difference in dB between the initial backscatter and the noise level at the longest pulse and a 3-minute average; about 6 dB of it are needed for a clean measurement, so the reachable length is roughly (dynamic range − 6 dB − sum of event losses) / attenuation (Formulas & calculations).

Why bidirectional

The backscatter coefficient differs from fibre to fibre — a mode field diameter difference within the ±0.4 µm tolerance changes it by several tenths of a dB. A single-direction trace therefore reports a splice between two such fibres as too good in one direction and too bad in the other, by up to ±0.3 dB. The true loss of every splice and connector is the average of the two directions, and TIA-568.3-D Tier 2 and the IEC 61280-4 series require bidirectional averaging when splice and connector losses are reported. In practice: an OTDR with a remote unit, or two acquisitions from the two ends, and software that matches the events and averages them.

From trace distance to route position

FactorEffect
Fibre length vs cable length0.1–0.5 % excess fibre in loose tubes; the OTDR measures fibre, the drawing shows cable
Cable length vs sheath marksuse the metre marks recorded at both ends of each cable in the closures (Documentation & labelling)
Slack loops10–30 m stored at every closure, more at poles and building entries
Aerial sag1–3 % longer than the pole-to-pole distance
Duct meanderingthe GIS line is shorter than the duct
IOR error0.1 % per 0.0015 of index

Method: find the nearest known event on the trace (a closure, a panel) and measure from it, not from the instrument; convert with the recorded metre marks; confirm with a trace from the other end; use a visible fault locator or clamp-on identifier for the last metres.

Standard misreadings

What you seeWrong conclusionRight reading
A spike beyond the fibre end"there is more fibre"a ghost of a strong reflector
A negative splice loss"perfect splice"a gainer — average both directions
No connector visible after the panel"no loss there"it is inside the attenuation dead zone — launch cord, shorter pulse
Large loss at 1550 nm only"bad splice"a macrobend; the splice is fine
High slope along a whole section"bad fibre"stress, water, ice or old fibre; the slope does not depend on IOR
Distances that disagree with the drawing by 3 %"records are wrong"excess fibre and slack loops — convert properly
Many small events near the noise"many splices"noise — average longer or use a longer pulse
The end found at 17 dB into a PON"fibre ends at the splitter"the end-detection threshold is too low for a splitter
Trace ends early with no spike"no break, just range"a wet or angled break — check from the other end
Connector reflectance reads −60 dB on a flat-topped spike"excellent connector"saturation — the value is meaningless

Live fibre and PON

On a live fibre an OTDR works at 1625 or 1650 nm through a filter that blocks the traffic wavelengths in both directions. In a PON, a trace from the ONU side upward shows the drop, the splitter and the feeder cleanly; a trace from the OLT side shows all branches superimposed after the splitter and cannot separate them. Splitters need a PON-optimised OTDR with high dynamic range and short dead zones, and the distance to each splitter must come from the records (PON problems).

Reporting

Keep the native .sor file (Telcordia SR-4731) for every fibre, direction and wavelength, with the pulse, range, IOR and launch-cord length stored in it; a PDF event table is not re-analysable. The report shows bidirectionally averaged event losses, reflectances, section attenuation and total loss against the project thresholds. Under TIA-568.3-D and ISO/IEC 14763-3 the OTDR trace is the Tier 2 characterization; pass/fail of the link loss is the Tier 1 light-source-and-power-meter test (Testing & measurement, Fibre characterization).

In CodingBox

The transceivers measure the same plant from the outside: far-end Tx power minus near-end Rx power is the plant loss including everything the OTDR hides in its dead zones — the module ports and the first and last connectors. When the DDM loss exceeds the OTDR total by more than a decibel, the difference sits in those hidden places, and a bench read of the module with a clean reference cord in CodingBox tells whether the module port or the panel is the dirty one (DDM in the app, Rx power & link budget).


If you find an inaccuracy or error in this material, select the relevant fragment and press Ctrl+Enter to .